BECKMAN COULTER
Power Express CENNEXUS Software
|
QUOTATION
INFORMATION
|
DESCRIPTION |
POWER EXPRESS |
DESCRIPTION
The Power Express system consists of modules configured to accommodate the needs of your laboratory. This section identifies the common hardware components and modules of the basic configuration along with the optional modules available for the Power Express. Your system may not be configured with all modules.
Houses the following software applications for Power Express:

Cennexus – Process control software for Power Express
- Software works with hardware to automate processing and manage patient sample tubes through preanalytical, analytical, and postanalytical processes
- Receives sample programming from the LIS and then downloads routing information to Line Control
- Bar code : Bar code reader reads primary sample tube bar code label to identify from Cennexus
how many secondary tubes the sample programming requires
- Aliquotter : Based on Cennexus setting, option is available to let the primary tube continue
processing on connected analyzers even though there may be insufficient volume to create all requested secondary tubes for all requested tests
- Cennexus is updated with reagent status and sample status for dynamic analyzer
connections. Samples going to non-dynamic connections are considered complete once they reach the module
Line Control – Controls sample movement through the entire system
- Sends and receives informationfrom Cennexus and the Power Express track system
- Provides status of connected modules
Power Express communications
- Cennexus receives sample programming information from the host (LIS or middleware) and routes it to Line Control
- Line Control communicates with Cennexus to route sample tubes on the Power Express system
- Cennexus provides dynamic interface to AU680, AU5800, DxC, and DxI analyzers.
- This two-way communication supports real-time updates to analyzer availability, test availability, sample tube arrival or rack load, and test completion
- Cennexus supports non-dynamic analyzers such as hematology, hemostasis (coagulation), and non-Beckman Coulter connected analyzers
CENNEXUS Software Area Description

1 Dashboard Displays across the top of each screen and includes the System Status and seven buttons. The buttons include Loading Pause, three Alerts, Audible On/Off, Login, and View Event Log
2 Main Menu Displays on the left panel of each screen and contains nine buttons used to select various system functions
3 Work Space Displays the information and functions unique to each Cennexus screen
3 Function Buttons Display in the work space and specific to the current screen
4 Notification Bar Displays notifications, user information, and the date and time at the bottom of the Cennexus screen
POWER EXPRESS
BECKMAN COULTER website
|
BECKMAN COULTER
Solutions and systems for clinical diagnostics (hematology, biochemistry, immunoassays, flow cytometry. Solutions for industry & research, for genetics, proteomics and cellular analysis. Lab automated workstation & liquid handling workstation.
|
Prices are quoted in euro (taxes excluded) and apply in the Benelux (recycling fee included). They are subject to change without notice. We reserve the right to adjust pricing errors and to limit quantities.
|